Rupture of Accura balloon during percutaneous transvenous mitral commissurotomy

Pradhan S, Sahoo PK, Mallick S, Padhi J K

Abstract


Percutaneous transvenous mitral commissurotomy (PTMC) is one of the treatment modality in the management of rheumatic mitral stenosis. Reuse of sterilized PTMC balloon catheters has been widely practiced in developing countries to bring down procedural cost. PTMC by reused balloons  are prone to deform and rupture during procedure. This can cause fatal complications like embolism of air or balloon material. Herein, we report a case of Accura balloon rupture during PTMC without any fatal complication. Thus, during balloon preparation particularly of reused balloon, it should be examined for any deformity or tear and air should be removed completely to prevent the fatal outcome.

Full Text:

PDF


DOI: https://doi.org/10.5430/crim.v4n1p1

Refbacks

  • There are currently no refbacks.


Case Reports in Internal Medicine

ISSN 2332-7243(Print)  ISSN 2332-7251(Online)

Copyright © Sciedu Press

To make sure that you can receive messages from us, please add the ‘sciedupress.com’ domains to your e-mail 'safe list'. If you do not receive e-mail in your 'inbox', please check your 'spam' or 'junk' folder.